Installing Dionex Devices
Dionex ESA Corona (plus/ultra/ultra
RS) CAD: Installation
Device Type: |
Charged Aerosol Detector |
Device Driver: |
Dionex ESA Modules: Corona Charged Aerosol Detector |
Hardware
Options |
The driver supports the following Corona models:
|
Restrictions: |
The following detector functions are not supported:
|
What is required?
Note:
If you run Agilent LCs controlled by Empower, you require the Corona DII connection kit (6081.2100) for connecting the Corona detector. Installation instructions are provided with the connection kit.
Hardware Installation
Safety Connection: |
Caution: Dionex recommends connecting the PUMP OFF output contact of the detector to the pump, so that, if the pressure of the nebulizing gas falls below 50% of the required gas volume, the signal may be used to shut the pump off so that the detector is not flooded with mobile phase. |
Device Connection: |
First, install the Chromeleon software, and then connect the instrument to the USB port on the Chromeleon server PC.
|
Software Installation
In the Chromeleon Server Configuration, add the driver to a timebase. For details, see Adding, Configuring or Deleting Devices. After you add the driver, the configuration dialog opens automatically.
On each configuration tab page, select the desired settings. For information about the settings on a page, click Help or press F1.
Further Information
For detailed installation instructions, refer to the operating instructions for the instrument.
For troubleshooting information, refer to Troubleshooting for USB and TCP/IP Connections.
For an overview of the different ESA devices, refer to ESA Overview.